- ClassificationCategoryCybersecurityDomainCybersecurityFramework FamilyISO Industry Standards
- Regulatory ContextTypeGuidanceLegal InstrumentTechnical ReportSectorHealthcare SectorIndustryHealthcare & Life Sciences
- Region / PublisherRegionGlobalRegion DetailInternationalPublisherInternational Electrotechnical Commission (IEC), TC 62/SC 62A
- VersioningVersionEdition 1.0 (2021-01); stability date 2028Effective DateJanuary 18, 2021Issue DateJanuary 18, 2021
- AdoptionAdoption ModelVoluntaryImplementation ComplexityMedium
- Official ReferenceOpen Link in New TabSource
License included / downloadable: No
IEC TR 60601-4-5:2021 is a paid publication. It must be purchased from the IEC Webstore or from a national standards body, and SmartSuite does not redistribute its text.
Framework text is licensed by its publisher and is included only where stated above.
- IEC TR 60601-4-5:2021 is an IEC Technical Report with security specifications for medical devices used in medical IT-networks.
- Edition 1.0 was published in January 2021 by IEC subcommittee 62A; it is guidance, not a certifiable standard.
- It is based on the seven foundational requirements of IEC TS 62443-1-1 and modifies IEC 62443-4-2 only for specific medical device aspects.
- Component requirements and requirement enhancements define a device's capability security level, SL-C, which integrators use to place the device in a zone.
- The report sets no target level for a whole network and does not apply to in-vitro diagnostic devices.
Overview
IEC TR 60601-4-5:2021 is a Technical Report from the International Electrotechnical Commission (IEC), Edition 1.0, published in January 2021. It gives detailed technical security specifications for medical devices used in medical IT-networks: medical electrical equipment, medical electrical systems and medical device software. Its main audience is medical device manufacturers and, where appropriate, compliance authorities. It is guidance, not a certifiable standard.
IEC subcommittee 62A prepared the report. It is based on the seven foundational requirements of IEC TS 62443-1-1:2009 and modifies IEC 62443-4-2:2019 only for specific aspects of medical devices. Component requirements (CRs) and requirement enhancements (REs) together define a device's capability security level (SL-C). Medical IT-network integrators use the SL-C to choose further countermeasures for the zone the device joins.
Manufacturers apply it by testing a device against the common security constraints in clause 4, deciding which specifications are native to the device and which the hospital network can provide, and documenting the SL-C and integration instructions (clause 7 and Annex A.6). The report sets no target level (SL-T) for a whole network and does not apply to in-vitro diagnostic devices.
Why does it matter?
The report lets a manufacturer state a device's security capabilities as a capability security level (SL-C), so a hospital or integrator can place the device in a zone of its medical IT-network without assessing it from scratch. Its focus is safety-related security: after a successful attack on the network, the device must still provide basic safety and essential performance (Introduction, essential function).
Key benefits include:
- A common level for device capability
The SL-C expresses how strong a device's countermeasures are for each of the seven foundational requirements, and it contributes to the achieved level of the part of the network the device sits in.
- Testable properties
Annex A.1 takes the approach of type-testable medical device IT security properties, so a claimed capability can be verified on a device rather than argued.
- Safety kept in view
Each proposed countermeasure should take into account that the safety and performance of the medical device must not be negatively affected.
- Flexibility for hospital networks
A device may not provide a capability itself but may integrate with a higher-level entity such as a hospital network. The report guides which specifications must be native to the device.
- Clear integration documentation
Manufacturers should document how to integrate the device into a medical IT-network, and Annex A.2 shows the typical network connections of medical devices.
How does it work?
The report starts from the seven foundational requirements defined in IEC TS 62443-1-1:2009: identification and authentication control (IAC), use control (UC), system integrity (SI), data confidentiality (DC), restricted data flow (RDF), timely response to events (TRE) and resource availability (RA). For medical devices it specifies component requirements (CRs), mainly derived from the system requirements in IEC 62443-3-3, and requirement enhancements (REs). The combination built into a device determines its SL-C.
Clause 4 sets common security constraints: support of the essential function (4.2), compensating countermeasures (4.3), least privilege (4.4), data minimization (4.5) and overarching constraints such as hardware security and specific security features for medical devices (4.6). Clause 5 explains how security levels apply to each foundational requirement, and clause 7 maps each single requirement to an SL-C in Table 1.
Countermeasures can be technical (firewalls, anti-virus software), administrative (policies and procedures) or physical (locked doors, encapsulated circuit boards). A manufacturer decides which specifications are native to the device and which can be allocated to the hospital or department network, then documents the SL-C and the integration instructions (Annex A.6).
Key Elements
- Seven foundational requirements
IAC, UC, SI, DC, RDF, TRE and RA, taken from IEC TS 62443-1-1:2009 and used to define the capability security level.
- Capability security level (SL-C)
A measure of the strength of countermeasures, separate or built in, for each security property of a device. Clause 5 and Table 1 in clause 7 map requirements to it.
- Component requirements and enhancements
CRs and REs for medical devices, mainly derived from IEC 62443-3-3 system requirements. Together they set the SL-C a device achieves.
- Common security constraints (clause 4)
Essential function, compensating countermeasures, least privilege, data minimization and overarching constraints including hardware security.
- Essential function and essential performance
The report separates the security term essential function from the safety term essential performance, and expects the device to sustain minimum functions, including basic safety, after an attack.
- Informative Annex A
Rationale and guidance: type-testable properties (A.1), network connections (A.2), zones and conduits with target levels (A.5), documentation of SL-C (A.6) and correlation with IEC TR 80001-2-2 (A.8).
What does it cover?
The report applies to medical devices with external data interfaces, such as a connection to a medical IT-network, or with a human interface used to enter, capture or view confidential data (clause 1). Medical devices include medical electrical equipment, medical electrical systems and medical device software. It does not apply to in-vitro diagnostic devices, industry protocols such as DICOM and HL7, or software on the network that is not part of a medical device.
What are its objectives?
The report is meant to help manufacturers build, and networks integrate, medical devices with defined security capabilities. It sets out to:
Specify security capabilities that let a medical device be integrated into a medical IT-network at a given security level.
Provide a flexible framework for addressing current and future vulnerabilities and applying mitigations in a systematic, defensible manner.
Give manufacturers specifications for each capability security level and show which must be native to the device.
Let network integrators identify a device's realized SL-C and specify additional countermeasures.
Protect the safety and performance of the device while security measures are applied.
Framework in Context
The report references IEC 62443-4-2:2019 and IEC 60601-1:2005 (with Amendments 1 and 2) as normative references (clause 2). Medical IT-networks as a whole, including the target level SL-T, are outside it and fall under standards such as IEC 80001 (all parts). Annex A.8 explains the correlation with IEC TR 80001-2-2. IEC itself provides no attestation of conformity to it.
Common Framework Mappings
Organizations map IEC TR 60601-4-5:2021 to other recognized frameworks to enhance interoperability, simplify compliance with industry standards, and strengthen risk management practices across medical device cybersecurity programs.
Mapped frameworks include:
IEC 62304
ISO 14971:2019
ISO 27001:2022
ISO 27002:2022
ISO/IEC 80001-1
MITRE ATT&CK
NIST CSF 2.0
NIST 800-53 Rev.5
How SmartSuite Supports IEC TR 60601-4-5:2021
SmartSuite provides the system for managing controls, evidence, mappings, assessments, and reporting.
Use Cyber & IT Risk to hold each device's SL-C by foundational requirement, the clause 4 constraints it meets and the integration documentation you give hospitals, each with an owner, evidence and a review date.
SL-C Record per Device
Record the capability security level of each device across the seven foundational requirements, with the CRs and REs it implements.
Common Constraints Checklist
Track the clause 4 constraints for each device: essential function, compensating countermeasures, least privilege, data minimization and hardware security.
Requirement Mapping to Table 1
Map each requirement in clause 7 to the device feature, the test evidence and the responsible engineer.
Native and Allocated Specifications
Log which specifications are native to the device and which are allocated to the hospital network, with the integration documents you issue.
Network Connection Profiles
Describe each device's typical network connections (Annex A.2) and the zone and target level SL-T an integrator assigns.
Gap Reporting and Review
Report open gaps against each SL-C to risk owners from the same records, with dashboards and review dates.
Related frameworks

ISO/IEC 27001:2022 specifies requirements for an information security management system: Clauses 4 to 10, risk treatment, the Statement of Applicability and 93 Annex A controls.

ISO/IEC 27002:2022 is the reference set of 93 information security controls in four themes, with attributes and implementation guidance, that supports an ISO/IEC 27001 ISMS.

ISO/IEC 27018:2025 gives guidelines for protecting PII in public clouds acting as PII processors, built on ISO/IEC 27002:2022 with an extended control set.
Frequently Asked Questions For IEC TR 60601-4-5:2021 (Medical Electrical Equipment Cybersecurity Guidance)
It is an IEC Technical Report, Edition 1.0, published in January 2021, titled Medical electrical equipment, Part 4-5: Guidance and interpretation, Safety-related technical security specifications. Clause 1 says it provides detailed technical specifications for the security features of medical devices used in medical IT-networks.
It is a Technical Report, so it is guidance and not a certifiable standard. The IEC foreword states that IEC itself provides no attestation of conformity. The introduction names medical device manufacturers as the main audience and compliance authorities as a further audience, so a regulator or customer may ask for it.
The introduction says the report references the security level requirements for network components in IEC 62443-4-2:2019 and modifies that standard only for specific aspects of medical devices in medical IT-networks. Its component requirements are mainly derived from the system requirements in IEC 62443-3-3.
The seven foundational requirements from IEC TS 62443-1-1:2009 are identification and authentication control, use control, system integrity, data confidentiality, restricted data flow, timely response to events and resource availability. The capability security level (SL-C) measures the strength of a device's countermeasures against them, and clause 7 maps requirements to it.
Clause 1 excludes in-vitro diagnostic devices, industry protocols such as DICOM and HL7, and software on a medical IT-network that is not medical device software. The introduction also leaves out the target level (SL-T) and achieved level (SL-A) of a whole network, which belong to standards for medical IT-networks such as IEC 80001.